Abstrait

RP-HPLC Method Development and Validation for Simultaneous Estimation of Thiocolchicoside and Lornoxicam in Tablet Dosage Form

B. Rajani and K. Mukkanti


A simple and cost effective reverse phase high performance liquid chromatography method was developed for simultaneous estimation of Thiocolchicoside and Lornoxicam in tablet dosage form. In this mehod, Chromatography was carried out on an Hypersil BDS, C-18 column (4.6 x 250 mm, 5 μ particle size) with a mobile phase composed of acetonitrile -phosphate buffer pH-3 in Gradient technique at a flow rate of 1.0 mL/min. Detection was carried out using a PDA detector at 260 nm. Validation parameters were studied as per ICH guidelines. The retention times for Thiocolchicoside and Lornoxicam are 2.3 min and 6.1 min, respectively. The linearity range for Thiocolchicoside and Lornoxicam are 10-60 µg/mL and 20-120 µg/Ml, respectively. The percentage recoveries of Thiocolchicoside and Lornoxicam are 99.96% and 100.65%, respectively. The correlation coefficients for both components are close to 1. This method is fast, accurate, precise and sensitive hence it can be employed for routine quality control of tablets containing both drugs in industries


Avertissement: testCe résumé a été traduit à l'aide d'outils d'intelligence artificielle et n'a pas encore été examiné ni vérifié

Indexé dans

  • CASS
  • Google Scholar
  • Ouvrir la porte J
  • Infrastructure nationale du savoir de Chine (CNKI)
  • CiterFactor
  • Cosmos SI
  • Bibliothèque de revues électroniques
  • Répertoire d’indexation des revues de recherche (DRJI)
  • Laboratoires secrets des moteurs de recherche
  • ICMJE

Voir plus

Flyer