New Design Units in Flow Injection Analysis and Sequential Injection Analysis for Determination of Copper (II) by Analytical Reagent

Khdeeja Jabbar Ali and Safa Ziyed Saleem


A simple and facial flow injection method coupled with spectrophotometer detection was developed for determination of copper(II) using flow injection analysis (FIA) and sequential analysis (SIA). The FIA and SIA systems based on the reaction of copper (II) with EDTA at pH 4-5 with an absorption maximum at 732 nm. This study was performed with different physical and chemical parameters such as flow rate, reaction coil, volume of reagent (EDTA), volume of sample, pH medium and concentration of EDTA. The calibration curve was prepared and the dispersion coefficient, reproducibility and application were studied. Two methods, flow injection and sequential injection analysis were compared for determination of copper (II) by EDTA. The linear range was (0.6-12 mg/L), (1-16 mg/L) at sampling rate of 137, 114 sample per hour, the detection limits (0.218 mg/L), (0.136 mg/L) for FIA and SIA, respectively. Relative standard deviations for (10 mg/L), n = 10 were found (0.577% for FIA and 0.45% for SIA). Dispersion coefficient was also measured for the two methods.